Facility

Source

Fluence-average neutron energy [MeV]

JAEA

Thermal [12]

0.025 x 10-6

JAEA

D2O moderated [14]

0.55

JAEA

Graphite moderated [18]

0.84

JAEA

Concrete moderated 3 [13]

0.58

JAEA

Concrete moderated 2 [13]

1.5

JAEA

Concrete moderated 1 [13]

2.6

AIST, JAEA

252Cf

2.13

JAEA

241Am-Be

4.16

Tohoku Univ.

Iron moderated [20]

0.33

Table 1: Fluence-averaged neutron energy of monoenergetic neutrons

Facility

Source

Neutron energy [MeV]

JAEA

Sc(p, n) [15]

0.008

AIST

Li(p, n) [11]

0.024

JAEA

Li(p, n) [16,17]

0.144

JAEA

Li(p, n) [16,17]

0.250

JAEA

Li(p, n) [16,17]

0.565

Tohoku Univ.

T(p, n) [19]

1

AIST

T(p, n) [10]

1.1

Tohoku Univ.

D(d, n) [19]

5.0

JAEA

T(d, n) [17]

14.8

Table 2: Fluence-averaged neutron energy of monoenergetic neutrons

Measurable energy range

0.025 eV to 20 MeV

Measurement range

Dose rate

0.1 μSv h-1 to 100 mSv h-1

Accumulate Dose

0.01 μSv to 100 mSv

Accuracy

Dose rate

≤ ±20 % at 10 μSv h-1

Accumulate Dose

≤ ±20 % at 1 μSv

Angular dependence

≤ ±25 %

Neutron sensitivity (252Cf)

NSN3A
NSN3B

0.3 s-1/(μSv h-1)
0.2 s-1/(μSv h-1)

Photon sensitivity (137Cs and 60Co)

1 mSv h-1
10 mSv h-1

< 0.01 %
< 0.1 %

Table 3: Specifications of neutron survey meter, NSN3
Figure 1: Cross-sectional views of the developed neutron survey meters
Figure 2: Minimum inhibitory concentration of methanolic extract (RME) of Rosa indica hip seeds for microbial strains (162) testedCalculated results of neutron energy response with changing the polyethylene thickness inserted between the detector and the shield box
Figure 3: Experimental arrangement at AIST, JAEA and Tohoku University
Figure 4: Neutron energy response for neutron fluence of the developed NSN3B neutron detector by comparing with the NSN3A neutron detector. Measured data are open circle, calculated data using PHITS code are solid line of histogram. Bold solid line is the fluence-to-ambient-dose-equivalent conversion coefficient, H*(10)/ϕ, given by ICRP Pub. 74
Figure 5: Neutron ambient dose equivalent response of the developed NSN3A and NSN3B neutron survey meters for various continuous energy neutron fields in very wide energy range from 0.025 eV up to about 300 MeV
Figure 6: Angular distribution of the developed neutron survey meter using an 241Am-Be neutron source
Figure 7: Outlook of light-weiht portable neutron survey meters, NSN3A and NSN3B

Active portable neutron survey meters that are widely used in nuclear facilities; however, they have generally heavy weight above 7 kg [1-5]. It is because in those conventional neutron survey meters, a central thermal neutron detector, such as a 3He counter, BF3 counter, and LiI(Tl) scintillator, is covered with thick (about 10 cm) polyethylene moderator. The handling of these neutron survey meters is therefore quite inconvenient for a person who walks around a large site to survey the neutron dose level by carrying the heavyweight survey meter. Among the three counters, the 3He counter has mostly been used as a central thermal neutron detector due to higher thermal neutron sensitivity [6,7]. However, the recent shortage of 3He gas, so-called 3He crisis, which has been a worldwide serious problem in neutron field measurement

Considering these circumferences, we have newly developed two types of lightweight neutron survey meters using a gas counter with a mixture of methane and nitrogen for dose management in nuclear power plants and accelerator facilities. These two survey meters, NSN3A and NSN3B, are approximately 16 cm x 25 cm x 30 cm in size, and can measure ambient neutron dose equivalent H*(10), directly from thermal to about 20 MeV neutrons. One type without polyethylene moderator, NSN3A, is approximately 2.2 kg in weight, which is 70% or more light weight than the conventional moderated-type neutron survey meters. The detailed information of this survey meter has recently been published in Ref. 8. The NSN3A survey meter has very good energy response to H*(10) only within 25% difference to the continuous energy neutron fields which are close to the actual workplace neutron fields. However, the response to the mono-energetic neutron fields in keV energy region is highly deviated to H*(10). This can bring the underestimation of the neutron dose equivalent such as in a passageway and in a personal airlock in nuclear power plant, which are dominated by low-energy neutrons below 100 keV.

Then we further improved the NSN3A survey meter with inserted thin polyethylene moderator of 2.5 cm thickness in order to increase the sensitivity of neutrons especially in the keV region. This new survey meter, NSN3B, has a little heavier weight of 3.5 kg, but still lighter than the conventional polyethylene moderated survey meters of about 7 to 11 kg. Here in this work, the further development of NSN3B is given by comparing it with NSN3A.

The characteristics of these two neutron survey meters for neutron energy response, angular dependence and detection efficiency were evaluated by measurements at several mono-energetic and quasi-mono-energetic neutron reference fields from thermal energy up to 244 MeV. The neutron energy responses of these two survey meters were also evaluated using the Particle and Heavy Ion Transport Code System (PHITS) code [9].


Outline of neutron detectors

The cross-sectional views of the two survey meters are shown in Figure 1 (a) and (b). The neutron detectors of both NSN3A and NSN3B survey meters are a gas counter consisted of methane gas of approximately 0.4 MPa and nitrogen gas of approximately 0.1 MPa which are encapsulated into thin stainless-steel vessel. It is almost a spherical shape of approximately 13-cm diameter and effective length of 13 cm. The effective volume is approximately 1440 cm3. The neutron detectors of NSN3A and NSN3B survey meters are both encapsulated into 1 mm thick aluminum shield box, and only for the neutron detector of NSN3B the thin polyethylene moderator of 2.5 cm thickness is inserted between the detector and the shield box.

Figure 1

The neutrons are measured using the mixed gas of methane and nitrogen. Fast neutrons above several hundreds keV are measured using the recoil proton of elastic scattering reaction of hydrogen, H(n, n)p, in the methane gas and slow neutrons including thermal neutrons are measured using the 14N(n, p)14C reaction of nitrogen gas. The proton energy in 14N(n, p)14C reaction is 626 keV. The electrical pulses are generated from those protons. By using these two reactions the neutron ambient dose equivalent can be obtained from thermal neutrons up to about 20 MeV neutrons. For higher energy region, carbon reactions such as 12C(n, p) and 12C(n, α) reactions, have some contribution to the output pulses.

Evaluation of neutron and gamma-ray characteristics

The neutron energy response and detection efficiency were measured in the continuous energy neutron reference fields and the monoenergetic neutron fields at the National Institute of Advanced Industrial Science and Technology (AIST) [10], the Facility of Radiation Standard (FRS) [11-17] of Japan Atomic Energy Agency (JAEA), the Fast Neutron Laboratory (FNL) [18] and the Cyclotron and Radioisotope Center (CYRIC) [19] of Tohoku University. The fluence-averaged neutron energies in various neutron reference fields are shown in Table 1 for continuous energy neutron fields and in Table 2 for monoenergetic neutron fields. A graphite pile of 150 cm x 164 cm x 150 cm loading a 252Cf source set at the center of graphite pile [11] is used for both thermal neutron field and graphite-moderated neutron field [17]. The graphite-moderated neutron field is produced by using the 241Am-Be source at the shallow position with the thermal neutron absorption sheet made of gadolinium. A concrete moderated neutron field is a neutron field (now defunct) where the 241Am-Be source is transmitted through concrete blocks of 1 m width, 2.85 m length and 1 m height [12]. The neutron detector is fixed at 53 cm height in space and the neutron spectrum can be changed by keeping 60, 110 and 160 cm distance between the source and the concrete blocks. These two neutron fields have wide-spread neutron spectra simulating workplace neutron fields.

Although these neutron survey meters are originally aimed to measure the ambient neutron dose equivalent up to 20 MeV, we further investigated the neutron energy response of the Type A survey meter for higher energy region above 100 MeV for use in high energy accelerator facilities. For this application, neutron detection efficiency was investigated by measurements at quasi-mono-energetic neutron reference fields of 134, 197 and 244 MeV produced from the 7Li(p, n) reaction at Research Center for Nuclear Physics (RCNP) cyclotron facility of Osaka University [20].
The influence of the gamma rays was measured using 137Cs and 60Co sources at 1 mSv h-1 and 10 mSv h-1 at Fuji Electric Co., Ltd.


Neutron energy responses were calculated using the Monte Carlo simulation, PHITS (Particle and Heavy Ion Transport code System) 2.64 [9] with JENDL 4.0 [21] neutron cross section data. In this calculation, carbon, hydrogen and nitrogen were installed in the thin stainless-steel vessel of which shapes was almost sphere and the energy deposition distributions from charged particles were obtained by T-Deposit tally named in PHITS using the neutron energy from 0.001 eV to 100 MeV. The neutron energy was divided by 10 energy regions in the log-scale at every digit and neutrons were generated more than 107 events in order to decrease the statistical errors. Neutron energy responses were obtained by summing up the counts, which were corrected using spectrum-weight G-function method described in Ref. 8 above the threshold level of gamma-ray cutoff of this detector to be decided at 400 keV.

Table 1

Table 2

Figure 2 gives the calculated results by changing the polyethylene thickness inserted between the detector and the shield box. The neutron energy responses were normalized at 2 MeV to obtain the relative neutron response. With increasing the polyethylene thickness, the calculated neutron energy response in the keV region largely increases and approaches to the fluence-to-ambient-dose-equivalent conversion coefficient, H*(10)/ϕ. We therefore decided to have 2.5-cm-thick polyethylene as described before.

Figure 3


Figure 3 shows the experimental arrangement. Signals from the neutron survey meter after the amplifier unit were fed into an external multi-channel analyzer, MCA600 (Laboratory Equipment Corp.) to obtain the pulse height distribution corrected using the G-function method [8]. The detector was located at the distance of 80 to 250 cm from the neutron source. In thermal neutron irradiation, the detector was located at a closer distance of 40 cm from the graphite pile surface. The room-scattered neutrons were subtracted with the shadow corn method, excluding the thermal neutron field, concrete-moderated neutron field and graphite-moderated neutron field.

Figure 3


Neutron energy response

The neutron energy responses of NSN3A and NSN3B for neutron fluence were obtained by summing up the counts which were corrected using the G-Function method above the discrimination level of 400 keV. The errors were estimated from statistical errors of the detector total counts, the neutron fluence at the reference field and the uncertainties of source intensities. Statistical errors of the total counts were usually less than 3.5 % (2σ : σ is the standard deviation), however, in the cases of 144 keV and 250 keV monoenergetic neutrons, the statistical errors were more than 95 %, due to the low neutron sensitivity in the energy around 100 keV. Uncertainties of neutron fluence of 252Cf, 241Am-Be sources and graphite-moderated neutrons were expected less than 14 % (2σ). Uncertainties of concrete-moderated neutrons were about 20 % (2σ). The thus-obtained neutron energy responses are indicated in Figure 4 by open circles for NSN3A and open square for NSN3B for comparing with the fluence-to-ambient-dose-equivalent conversion coefficient curve indicated by a bold solid line, H*(10)/ϕ, given by ICRP Pub. 74. The solid line and dashed line in histogram are the calculated results using the PHITS Monte Carlo code. The measured results are in good agreement with the PHITS calculation. The measured response of NSN3B becomes much better agreement with the fluence-to-ambient-dose-equivalent conversion coefficient curve indicated by a bold solid line, H*(10)/ϕ, about two orders of magnitude higher than that of NSN3A, which verifies much better neutron energy response in keV energy region.

The neutron energy responses at thermal energy and between 1 MeV and 10 MeV give good agreement with H*(10)/ϕ for both neutron survey meters, but the high-energy neutron response at 14.8 MeV gives underestimated values of approximately 40 % to H*(10)/ϕ. This is because that the neutron elastic scattering reaction cross section with the hydrogen decreases and the energy deposition of recoil protons decreases, due to decrease of the stopping power of proton with energy, then leads to decrease of the signals which can exceed the discrimination level of 400 keV.

The neutron ambient dose equivalent responses of NSN3A and NSN3B measured for various continuous energy neutron fields given in Table 1 and quasi-mono-energetic high energy neutron reference fields of 134, 197 and 244 MeV at RCNP [20,22], are shown in Figure 5, as response values normalized to that for 252Cf source as a function of fluence-averaged neutron energy. As a result, both NSN3A and NSN3B survey meters have sufficiently good response, within 35 % difference to neutron ambient dose equivalent for fluence-averaged continuous neutron energy in very wide energy range from 0.025 eV up to about 300 MeV. On the other hand, Fuji Electric NSN1, NSN2 [1] and Alnor 2202D which are both polyethylene moderated-type give about one-fifth lower values, while on the other hand, WENDI, which is polyethylene and tungsten moderated-type for higher energy use, gives about 30%–60% higher values at neutron energies over 100 MeV. It was found that NSN3A and NSN3B have much better energy responses than the conventional moderated-type neutron survey meters to be used at accelerator facilities of energy beyond 100 MeV.

Figure 4

Figure 5

Angular distribution

Angular distribution of NSN3A and NSN3B were obtained using an 241Am-Be neutron source at every 30-degree from 0 to 180 degrees in the vertical and horizontal directions at Fuji Electric Co. Ltd. The distance between the survey meter and the neutron source was 40 cm. Figure 6 shows the angular distributions of developed two neutron survey meters, the open circle for NSN3A and open square for NSN3B. These are normalized at zero degree of which the direction is shown in Figure 3. The angular distributions are almost constant within 25%, because the sensitive area of these detectors are almost a spherical shape as shown in Figure 1. It is found that these survey meters have a flat angular distribution for neutrons in all directions.

Figure 6

Figure 7

Response of gamma ray

The gamma ray responses of NSN3A and NSN3B were obtained using 137Cs and 60Co sources. These are under 0.01 % at 1 mSv h-1 and under 0.1 % at 10 mSv h-1. It is found that the developed neutron survey meters have almost very low sensitivity for photon radiation of dose rate below 10 mSv h-1.


The specifications of these two neutron survey meters, NSN3A and NSN3B, are shown in Table 3. The NSN3A survey meter without polyethylene moderator is only 2.2 kg in weight and the NSN3B survey meter with thin polyethylene moderator is 3.5 kg in weight. The NSN3A is over 70 % and the NSN3B is about half lighter than the conventional moderated-type survey meter, NSN1 and NSN2 [1], which are about 7 kg. These survey meters shown in Figure 7 are capable directly to measure ambient neutron dose equivalent H*(10) from thermal to about 20 MeV neutrons. The NSN3A has rather lower sensitivity to neutrons in keV region but has good energy response for continuous energy neutron field simulating actual wok-place neutron fields up to about 250 MeV neutron energy. The NSN3A survey meter is clearly a useful monitoring tool to use in various accelerator facilities over wide energy range. On the other hand, the NSN3B survey meter has much better energy response especially in keV region and is especially very useful to use in nuclear power plants where neutrons in the keV region occupy largely the neutron ambient dose equivalent rate.

Table 3


The two types of lightweight portable neutron survey meters have been developed comprising a proportional gas counter with a mixture of methane and nitrogen, the one without and the other with thin polyethylene moderator. Performance tests were carried out using mono-energetic and quasi-mono-energetic neutron fields from thermal energy up to about 250 MeV energy and also continuous energy neutron fields simulating actual work-place neutron fields. The neutron energy responses were obtained from thermal energy to fluence-averaged neutron energy of 4.16 MeV within 25 % difference to neutron ambient dose equivalent. Neutron detection efficiencies for 252Cf source were obtained to be 0.3 s-1/(μSv h-1) for NSN3A and 0.2 s-1/(μSv h-1) for NSN3B. These neutron survey meters have good response to ambient dose equivalent for continuous energy neutron sources simulating workplace and its surrounding environmental neutron fields. The NSN3A survey meter has also much better sensitivity up to about 250 MeV neutrons than the conventional moderated-type neutron survey meters, which becomes a useful monitoring tool in high energy accelerator facilities. The NSN3B survey meter has much better energy response especially in keV region and is especially very useful to use in nuclear power plants where neutrons in the keV region occupy dominantly the neutron ambient dose equivalent rate. Furthermore, these neutron survey meters have very low sensitivity to photons, less than 0.1 % for 10 mSv h-1 dose rates.

These survey meters satisfy the angler distribution, energy response and response to photon radiation in IEC 61005 standard [23]. The neutron detection efficiencies of 0.3 and 0.2 s-1/(μSv h-1) of these light-weight NSN3A and NSN3B survey meters are almost the same efficiency of Alnor 2202D having about 10 kg weight, although they are lower than 4.0 s-1/(μSv h-1) of high sensitivity survey meters, NSN1 and NSN2 having 7 kg weight. It can be finally concluded that these light-weight portable neutron survey meters of 2.2 kg and 3.5 kg can make the handling of neutron survey meter easier and be suitable for environmental monitoring around nuclear power plants and accelerator facilities.


The authors wish to thank the staff members of the AIST, JAEA and Tohoku University for their great support during the neutron irradiation


  1. Nakamura T, Hara A and Suzuki T (1985) Realization of a high sensitivity neutron rem counter. Nuclear Instruments and Methods in Physics and Research A241: 554-560.
  2. Birattari C, Esposito A, Ferrari A, Pelliccioni M and Silari M (1992) A neutron survey meter with sensitivity extended up to 400 MeV. Radiation Protection Dosimetry 44: 193-197.
  3. Bartlett DT, Tanner R and Jones DG (1997) A new design of neutron dose equivalent survey instrument. Radiation Protection Dosimetry 74: 267-271.
  4. Burgkhardt B, Fieg G, Klett A, Plewnia A and Siebert BRL (1997) The neutron fluence and H*(10) response of the new LB 6411 rem counter. Rad Prot Dosimet 70: 361-364.
  5. Klett A and Burgkhardt B (1997) The new remcounter LB6411: measurement of neutron ambient dose equivalent H*(10) according to ICRP60 with high sensitivity. IEEE Nuclear Science 44: 757-759.
  6. Kouzes RT (2009) The 3He Supply Problem, PNNL-18388.
  7. Kouzes RT, Ely JH, Erikson LE, Kernan WJ, Lintereur AT, et al. (2010) Woodring, Nutron detection alternatives to 3He for national security applications. Nucl Instr Meth Phy Res A 623: 1035-1045.
  8. Nunomiya T, Nakamura T, Yamamura S and Amano T (2017) Development of lightweight portable neutron survey meter. J Nucl Sci Tech 54: 1215-1222.
  9. Sato T, Iwamoto Y, Hashimoto S, Ogawa T, Furuta T, et al. (2018) Features of Particle and Heavy Ion Transport code System (PHITS) version 3.02.S J Nuc Sci Tech 55: 684-690.
  10. Harano H, Matsumoto T, Tanimura Y, Shikaze Y, Baba M, Nakamura T (2010) Monoenergetic and quasi-monoenergetic neutron reference field in Japan. Radiat Measurement 45: 1076-1082.
  11. Uchita Y, Saegusa J, Kajimoto Y, Tanimura Y, Shimizu S, Yoshiza¬wa M (2005) Characteristics of Thermal Neutron Calibration Fields Using a Graphite Pile. JAERI-Tech 012: 1-31.
  12. Yoshizawa M, Tanimura Y, Saegusa J, Nemoto H, Yoshida M (2004) Neutron spectra and angular distributions of concrete-moder¬ated neutron calibration fields at JAERI. Radiat Protect Dosimet110: 81-84.
  13. Kowatari M, Fujii K, Takahashi M, Yoshizawa M, Shimizu S, Kawasaki K and Yamaguchi Y (2007) Evaluation of the characteristics of the neutron reference field using D2O-moderated 252Cf source. Radiat Protect Dosimet126: 138-144.
  14. Tanimura Y, Saegusa J, Shikaze Y, Tsutsumi M, Shimizu S and Yoshizawa M (2007) Construction of monoenergetic neutron calibration fields using 45Sc(p, n)45Ti reaction at JAEA. Radiat Protect Dosimet 126: 8-12.
  15. Tanimura Y, Yoshizawa M, Saegusa J, Fujii K, Shimizu S, et al. (2004) Construction of 144, 565 keV and 5.0 MeV monoenergetic neutron calibration fields at JAERI. Radiat Protect Dosimet 110: 85-89.
  16. Shikaze Y, Tanimura Y, Saegusa J, Tsutsumi M, Shimizu S, et al. (2011) Development of the neutron calibration fields using accelerators at FRS and TIARA of JAEA. J Nuc Sci Tech 48: 1-30, 5: 209-212.
  17. Nishino S, Tanimura Y, Ebata Y and Yoshizawa M (2016) Development of the Graphite-Moderated Neutron Calibration Fields using 241Am-Be sources in JAEA-FRS. J Radiat Protect Res 41: 211-215.
  18. Baba M, Takada M, Iwasaki T, Matsuyama S, Nakamura T, et al. (1996) Development of monoenergetic neutron calibration fields between 8 keV and 15 MeV. Nucl Instru Meth Phy Res A376: 15-123.
  19. Nakamura T, Ohkubo T, Uwamino Y, Iwai S and Hara A (1986) Development of simple neutron reference calibration field and its evaluation. INS-T-453, Tokyo University in Japanese.
  20. Iwamoto Y, Hagiwara M, Satoh D, Araki S, Yashima H, et al. (2011) Characterization of high-energy quasi-monoenergetic neutron energy spectra and ambient dose equivalents of 80-389 MeV 7Li(p,n) reactions using a time-of-flight method. Nucl Instrum Meth Phy Res A804: 50-58.
  21. Shibata K, Iwamoto O, Nakagawa T, Iwamoto N, Ichihara A, et al. (2011) JENDL-4.0: A New Library for Nuclear Science and Engineering. J Nuc Sci Tech 48: 1-30.
  22. Nakane Y, Hagiwara M, Iwamoto Y, Iwase H, Satoh D, et al. (2014) Response measurement of various neutron dose equivalent monitors in 134-387 MeV neutron fields. Prog Nuc Sci Tec 4: 704-708.
  23. IEC 61005 (2014) Radiation protection instrumentation - Neutron ambient dose equivalent (rate) meters, International Electro-technical Commission 07.